Dräger Resuscitaire WBR82 Instructions For Use Manual page 122

Radiant warmer product
Hide thumbs Also See for Resuscitaire WBR82:
Table of Contents

Advertisement

PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04
Symptom
Heater does not switch on.
Pre-Warm Mode is not activated
upon a system reset.
Heater is ON, but the indicators
are OFF.
Cannot obtain the desired suction
pressure.
Blender O
concentration setting
2
does not match the blender set-
ting.
Blender output mixture is outside
the specified range (20.8% to
100%).
Patient Gas Supply measured
flow is outside of the specified
± 3% of full scale or 10% of set-
ting.
Patient Gas Supply pressure
relief gauge exceeds 50 cmH
± 10 cmH
O (4.9 kPa ± 1 kPa).
2
AutoBreath™ Infant Resuscita-
tor measured BPM is outside of
the specified range.
7 - 2
Possible Cause
Oxygen or air pressures are low.
Both cylinders may not be
opened.
Pressure differential between the
oxygen and Air supplies exceed
207 kPa (30 psi) ± 14 kPa (2 psi).
Heating element is open.
There is an internal malfunction. Refer the unit to service.
System is in Diagnostic Mode.
There is an internal malfunction. Replace the power and control
There is an internal malfunction. Replace the display P.C. board.
Suction jar rim gasket is worn.
Suction jar is loose or chipped.
Connecting hose is bent or tan-
gled.
Hose is leaking between the reg-
ulator and the bottle.
Pressure or suction gauge is not
measuring correctly.
Blender knob is loose.
Blender has failed.
Air inlet H
O filter is missing or
2
clogged.
Blender has failed.
Dirty or clogged air/O
Pressure relief valve has failed.
O
2
There is an internal malfunction. Refer the unit to service.
Corrective Action
Adjust oxygen or air pressure to
correct setting.
Check cylinder gauge.
Adjust oxygen or air pressure to
correct setting.
Refer the unit to service.
Exit Diagnostic Mode by press-
ing the reset button.
P.C. board.
Replace the suction jar.
Tighten or replace the suction jar.
Replace the hose.
Replace the hose.
Refer the unit to service.
Refer the unit to service. Do not
attempt to tighten the knob as
this voids unit calibration.
Replace the blender or refer the
unit to service.
Replace or install filter.
Replace blender or refer the unit
to service.
hose.
Replace the hose.
2
Replace the pressure relief valve
or refer the unit to service.

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Resuscitaire rw82Resuscitaire wmrw82

Table of Contents