PROPRIETARY AND CONFIDENTIAL DRAFT 8 Jan 04
WARNING:
During a gas failure, the patient only gets room air from the two-way relief valve on the Patient
Supply Resuscitation Module. Breathing room air through the two-way relief valve requires
extra effort. If this condition occurs, correct the gas failure as soon as possible. Failure to do so
could result in patient injury.
WARNING:
If the patient needs gas, ensure that the Gas Delivery Module gas flow is switched on. Failure to
do so could result in patient injury.
WARNING:
When setting the Flow Rate (LPM) control for optimum repeatability, turn the knob in a
clockwise direction to approach the desired setting. Failure to do so could result in patient injury.
WARNING:
Periodically measure blood gases to ensure proper levels of ventilation. Failure to do so could
result in personal injury.
WARNING:
Before use, confirm that the oxygen/air blender control of the blended gas supply module is
correctly set. Failure to do so could result in patient injury.
WARNING:
Ensure that the breathing circuit connections are secure and free of obstructions. Failure to do
so could result in patient injury.
WARNING:
Ensure that gas supplies are clean and dry. Use Water/Trap filters in the supply lines if
necessary. Replace according to the manufacturer's labeling. Failure to do so could result in
patient injury.
WARNING:
Securely fasten all hoses to fittings. Hand-tighten to avoid damaging the fittings. Failure to do so
could result in patient injury or equipment damage.
WARNING:
Do not use antistatic or electrically conductive hoses or tubing. Patient injury could occur.
3 - 9