Replace the Flow Cell / Cuvette Holder

Replace the Flow Cell / Cuvette Holder

When
If application needs a different type of flow cell or the flow cell needs repair.
Tools required
p/n
OR
5043-0915
Parts required
#
1
For flow cell details see
Preparations
Turn the lamp OFF.
1
Open the doors.
Agilent InfinityLab LC Series VWD User Manual
Description
Wrench, 1/4 inch
for capillary connections
Fitting mounting tool
Description
Flow cell
"Standard Flow Cell 10 mm / 14 µl"
"Micro Flow Cell 3 mm / 2 µl"
"Semi-micro Flow Cell (Parts)"
"High Pressure Flow Cell (Parts)"
on page 150
on page 152
on page 154
on page 156
2
Disconnect the inlet capillary and outlet tubing from the
flow cell.
1.
2.
135