Physical Specifications - Agilent Technologies InfinityLab LC Series User Manual

Variable wavelength detectors
Hide thumbs Also See for InfinityLab LC Series:
Table of Contents

Advertisement

2
Site Requirements and Specifications

Physical Specifications

Physical Specifications

Table 2
Type
Weight
Dimensions
(height × width × depth)
Line voltage
Line frequency
Power consumption
Ambient operating tempera-
ture
Ambient non-operating tem-
perature
Humidity
Operating altitude
Non-operating altitude
Safety standards:
IEC, EN, CSA, UL
ISM Classification
Agilent InfinityLab LC Series VWD User Manual
Physical Specifications
Specification
11 kg (24.3 lbs)
140 x 396 x 436 mm
(5.5 x 15.6 x 17.2 inches)
100 – 240 V~, ± 10 %
50 or 60 Hz, ± 5 %
80 VA, 70 W
4 - 55 °C (39 - 131 °F)
-40 – 70 °C (-40 – 158 °F)
< 95 % r.h. at 40 °C (104 °F)
Up to 3000 m (9842 ft)
Up to 4600 m (15092 ft)
Installation category II, Pollution degree 2
ISM Group 1 Class B
Comments
Wide-ranging capability
Non-condensing
For storing the module
For indoor use only.
According to CISPR 11
29

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

1290 infinity ii1260 infinity iiG7114bG7114a

Table of Contents