Error Information
Detector Error Messages
Detector Error Messages
These errors are detector specific.
UV lamp: no current
Error ID: 7450
The lamp anode current is missing. The processor continually monitors the
anode current drawn by the lamp during operation. If the anode current falls
below the lower current limit, the error message is generated.
Probable cause
1
2
3
4
5
Agilent InfinityLab LC Series VWD User Manual
Lamp disconnected.
Top foam removed while lamp is on.
Defective or non-Agilent lamp.
Defective main board.
Defective power supply.
Suggested actions
Ensure the lamp connector is seated firmly.
Please contact your Agilent service representa-
tive.
Exchange the lamp.
Please contact your Agilent service representa-
tive.
Please contact your Agilent service representa-
tive.