With Telnet - Agilent Technologies InfinityLab LC Series User Manual

Variable wavelength detectors
Hide thumbs Also See for InfinityLab LC Series:
Table of Contents

Advertisement

13
LAN Configuration
Manual Configuration

With Telnet

Whenever a TCP/IP connection to the module is possible (TCP/IP parameters set
by any method), the parameters may be altered by opening a Telnet session.
1 Open the system (DOS) prompt window by clicking on Windows START
2 Type the following at the system (DOS) prompt:
Agilent InfinityLab LC Series VWD User Manual
button and select "Run...". Type "cmd" and press OK.
c:\>telnet <IP address> or
c:\>telnet <host name>
Figure 61
Telnet - Starting a session
where <IP address> may be the assigned address from a Bootp cycle, a
configuration session with the Handheld Controller, or the default IP address
(see
"Configuration Switch"
When the connection was established successfully, the module responds
with the following:
Figure 62
A connection to the module is made
on page 196).