Leak; Leak Sensor Open - Agilent Technologies InfinityLab LC Series User Manual

Variable wavelength detectors
Hide thumbs Also See for InfinityLab LC Series:
Table of Contents

Advertisement

7
Error Information
General Error Messages

Leak

Error ID: 0064
A leak was detected in the module.
The signals from the two temperature sensors (leak sensor and board-mounted
temperature-compensation sensor) are used by the leak algorithm to determine
whether a leak is present. When a leak occurs, the leak sensor is cooled by the
solvent. This changes the resistance of the leak sensor which is sensed by the
leak-sensor circuit on the main board.
Probable cause
1
2
3

Leak Sensor Open

Error ID: 0083
The leak sensor in the module has failed (open circuit).
The current through the leak sensor is dependent on temperature. A leak is
detected when solvent cools the leak sensor, causing the leak-sensor current to
change within defined limits. If the current falls outside the lower limit, the error
message is generated.
Probable cause
1
2
3
4
Agilent InfinityLab LC Series VWD User Manual
Loose fittings.
Broken capillary.
Leaking flow cell.
Leak sensor not connected to the Power
Switch board.
Defective leak sensor.
Leak sensor incorrectly routed, being
pinched by a metal component.
Power switch assembly defective
Suggested actions
Ensure all fittings are tight.
Exchange defective capillaries.
Exchange flow cell components.
Suggested actions
Please contact your Agilent service representa-
tive.
Please contact your Agilent service representa-
tive.
Please contact your Agilent service representa-
tive.
Please contact your Agilent service representa-
tive.
85