10. System Checkout Procedures
This chapter details the test procedures that should be performed by
the clinician prior to connecting the device to the patient. Test the
Active PAP, Active Flow, and Passive circuit types if you want to do a
complete checkout on the device. The tests should be performed as
described in order to verify proper operation of the device. Some of
the procedures in this chapter require you to change settings on the
device. If you are not familiar with the procedures for accomplishing
this, please refer to Chapter 4, Ventilator Setup and Chapter 5, Viewing
and Changing Settings.

Tools Required

Universal Porting Block
Active PAP Exhalation Device
Active Flow Exhalation Device with Flow Sensor
Trilogy Universal Active PAP Tube Adapter
Passive Exhalation Device
Test Lung
High Pressure O
Hose
2
Oxygen Monitor
Small Flat Head Screwdriver
Trilogy 202
clinical manual
Note: The actual circuit
configuration to be used on
the patient should be used to
perform the system checkout
procedure.
WARNING
If you notice any unexplained
changes in the performance
of the device, if it is making
unusual sounds, if the device
or detachable battery are
dropped, if water is spilled into
the enclosure, or if the enclosure
is cracked or broken, discontinue
use and contact Philips
Respironics or an authorized
service center for service.
Chapter 10 System Checkout Procedures
137