2. Connect the other end of the flexible tubing to the active
exhalation device.
A. Connect the active exhalation device to the flexible
tubing that attaches to the ventilator.
B. Connect the flow sensor to the patient interface (e.g.,
the tracheostomy tube).
C. Connect the flow lines and the exhalation valve line to
the Universal Porting Block on the ventilator as shown
below.
Connect the White Striped Flow Line to
the top, striped port on the Universal
Porting Block on the device.
Connect the other Flow Line to the
middle port on the Universal porting
Block.
Connect the Exhalation Valve Line to the
Exhalation Valve port on the top of the
active exhalation device, and connect the
other end of the line to the Exhalation
Valve port on the Universal Porting Block.
White Striped
Exhalation Valve Port
Flow Line
Flow Line
Exhalation Valve Line
Refer to the instructions included with the active exhalation
device for more information.
Universal Porting Block
on Trilogy Device
Exhalation Valve Port
on Porting Block
Note: If the active exhalation
device has the proximal pressure
line connected when you are
setting up an active flow circuit
type, remove the proximal
pressure line and cap the
proximal pressure port before
attaching to the flexible tubing.
Note: Make sure that
components marked with an
arrow are oriented properly.
Note: When using an active
exhalation device with proximal
flow, ensure that any additional
proximal ports are capped.
Note: The
symbol
appears next to the Exhalation
Valve port on the Universal
Porting Block to indicate where
the active exhalation valve line
connects.
Connecting the Active
Exhalation Device with
Proximal Flow Sensor
Chapter 4 Ventilator Setup