144
Setting
Flow Pattern
PEEP
Sigh
Circuit Disconnect
Low Inspiratory Pressure
High Inspiratory Pressure
Apnea
All other alarms
B. Verify the Alarm
Wait up to 40 seconds and verify the following alarm signals:
The Medium Priority audible indicator sounds
A yellow light flashes on the
Pause button
The High Inspiratory Pressure alarm condition
appears on the screen, highlighted in yellow
C. Modify Ventilator Alarm Settings
Modify the High Inspiratory Pressure setting to match the one
shown below in Table 12.
Setting
High Inspiratory Pressure
D. Verify Reset
Wait 40 seconds and verify the following auto-reset conditions:
The Medium Priority audible indicator has stopped
sounding
The yellow light on the
button has stopped flashing
Trilogy 202 clinical manual
Value
Ramp
4 cmH
O
2
Off
Off
6 cmH
O
2
10 cmH
O
2
Off
Off
Alarm Indicator/Audio
Value
60 cmH
O
2
Alarm Indicator/Audio Pause
Table 12 - Modify High
Inspiratory Pressure Alarm
Setting