PHILIPS Respironics Trilogy 202 User Manual page 80

Ventilator
Table of Contents

Advertisement

If Detailed View is turned on in the Options menu, the Monitor
screen will look like the screen shown below.
This screen contains more detailed information about the
therapy.
The top Monitor panel contains the Prescription indicator
if a dual prescription exists, the therapy mode, a graph
displaying the current pressure, and the current date and
time. Additionally, this panel also displays patient pressure,
respiratory rate (RR), exhaled tidal volume (Vte), and leak.
The second panel in Detailed view is the Measured Settings
panel. It provides patient-related data including Peak
Inspiratory Pressure (PIP), I:E Ratio, Peak Inspiratory Flow,
Mean Airway Pressure (MAP), FiO
The third panel is the Status panel and shows the same
information displayed in the Detailed View Off screen,
including features in use such as Ramp and battery status.
Monitor Panel
Measured Settings
Panel
Status Panel
, and Minute Ventilation.
2
Chapter 5 Viewing and Changing Settings
Monitor Screen – Detailed
View On
71

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents