Pressure Control (PC) Mode
In Pressure Control (PC) mode, the device delivers bi-level pressure
support. This mode delivers Assist and Mandatory breaths. This
mode is identical to S/T mode, except that all breaths have a fixed
inspiratory time.
PRESSURE
6
seconds
26
20
10
3
6
Enter PC mode
5
Pressure Control - Synchronized Intermittent Mandatory
Ventilation (PC-SIMV) Mode
The Pressure Control – Synchronized Intermittent Mandatory
Ventilation (PC-SIMV) mode provides Spontaneous, Assist, and
Mandatory breaths. This mode uses a time window to decide
what type of breaths should be delivered. This time window is the
maximum time between breaths according to the set Breath Rate.
When you enter PC-SIMV mode, the time window is started. If
the patient does not trigger a breath, the ventilator will provide
a Mandatory breath when the time expires and then begin the
process again. Mandatory breaths deliver the Pressure setting during
inhalation and a lower Peak End Expiratory Pressure (PEEP)
during exhalation.
IPAP = 26 cm H
O
2
EPAP = 6 cm H
O
2
Rate = 10 BPM
PS = IPAP- EPAP = 20 cm H
O
2
6
6
seconds
seconds
3
2
10
15
= Mandatory Breath
2
= Assist Breath
3
3
20
TIME
Chapter 3 Modes, Features, and Alarms