PHILIPS Respironics Trilogy 202 User Manual page 35

Ventilator
Table of Contents

Advertisement

26
During the time window, if patient effort is detected, either a
Spontaneous or an Assist breath is delivered for the first effort. If the
last breath delivered was Mandatory, then a Spontaneous breath is
delivered for the first patient effort of the window. If the last breath
delivered was an Assist or Spontaneous breath, then an Assist breath
is delivered for the first patient effort of the window. Assist breaths
deliver the Pressure setting during inhalation and the PEEP setting
during exhalation.
During the time window, if patient effort continues to be detected
(after the first patient triggered breath), Spontaneous breaths are
delivered for the remainder of the window. The figure below provides
example breath patterns in PC-SIMV mode.
Pressure = 26 cm H
Pressure Support = 10 cm H
PEEP = 6 cm H
O
2
Rate = 5 BPM
PRESSURE
12 seconds
M
S
26
20
10
2
1
1
6
Enter PC-SIMV mode
5
10
Trilogy 202 user manual
O
= Spontaneous
1
2
O
Breath
2
= Mandatory Breath
2
= Assist Breath
3
12 seconds
12 seconds
M
S
3
1
15
20
25
30
= Mandatory Window
M
= Spontaneous Window
S
12 seconds
M
M
2
35
40
45
TIME
Sample Breath Patterns in
PC-SIMV Mode

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents