PHILIPS Respironics Trilogy 202 User Manual page 127

Ventilator
Table of Contents

Advertisement

118
Alarm
Priority
High Internal
High
Oxygen
High Oxygen
High
Flow
Low Oxygen
High
Flow
High Oxygen
High
Inlet Pressure
Trilogy 202 clinical manual
Audible
Visual (Alarm
Indicator Button
and Display)
• • • • •
Red flashing button;
"High Internal
Oxygen" message
• • • • •
Red flashing button;
"High Oxygen Flow"
message
• • • • •
Red flashing button;
"Low Oxygen Flow"
message
• • • • •
Red flashing button;
"High Oxygen Inlet
Pressure" message
Device
Clinician Action
Action
Continues
Disconnect supplemental
to operate
oxygen from device. Check
when internal
external oxygen connection.
oxygen
If problem continues, have
concentration
device serviced.
reaches
5% above
ambient
levels.
Continues to
Check hose from oxygen
operate
source to device. Check
oxygen source. The inlet
supply should be 175
SLPM. If problem continues,
contact Philips Respironics
or an authorized service
representative. Please
have the the device model
number and serial number
information when you call.
Continues to
Check hose from oxygen
operate
source to device. Check
oxygen source. The inlet
supply should be 175
SLPM. If problem continues,
contact Philips Respironics
or an authorized service
representative. Please
have the the device model
number and serial number
information when you call.
Continues to
Check oxygen source. The
operate
oxygen input must be
40-87 psi. If problem
continues, contact Philips
Respironics or an authorized
service representative. Please
have the the device model
number and serial number
information when you call.