154
Measured Patient Parameters
Parameter
Vte/Vti
Minute Ventilation
Estimated Leak Rate
Respiratory Rate
Peak Inspiratory Flow
Peak Inspiratory Pressure
Mean Airway Pressure
% Patient Triggered Breaths
I:E Ratio
All flows and volumes are measured at BTPS conditions.
Spontaneous Breathing During Power Failure Conditions
Flow Set
Point (l/min)
Active
Circuit
30
60
Disposal
Separate collection for electrical and electronic equipment per EC Directive 2002/96/EC. Dispose of
this device in accordance with local regulations.
Trilogy 202 clinical manual
Range
0 to 2000 ml
0 to 99 l/min
0 to 200 l/min
0 to 80 BPM
0 to 200 l/min
0 to 99 cmH
0 to 99 cmH
0 to 100%
9.9- 1: 1-9.9
Inspiratory Resistance (cmH
Active Circuit with
Proximal Flow
Sensor
< 3.0
< 3.5
< 9.0
< 10.5
Accuracy
Greater of 15ml or 15% of reading
Calculation based on measured Vte or Vti
and Respiratory Rate
N/A
Greater of 1 BPM or 10% of reading
3 l/min plus 15% of reading
O
Greater than 2 cmH
2
O
Greater than 2 cmH
2
N/A
Calculation based on Inspiratory Time and
Expiratory Time
O)
Expiratory Resistance (cmH
2
Passive
Active
Active Circuit with
Circuit
Circuit
< 2.0
< 2.0
< 5.0
< 3.0
O or 10% of reading
2
O or 10% of reading
2
O)
2
Passive
Proximal Flow
Circuit
Sensor
< 2.0
< 1.5
< 4.5
< 4.0