PHILIPS Respironics Trilogy 202 User Manual page 148

Ventilator
Table of Contents

Advertisement

Setting
Therapy Mode
AVAPS (passive circuit only)
IPAP
EPAP
FiO
2
Breath Rate
Inspiratory Time
Trigger Type (passive circuit
type)
Flow Trigger Sensitivity
(Active PAP or Active Flow
circuit type)
Leak Compensation (Active
Flow circuit type)
Flow Cycle Sensitivity
(Active PAP or Active Flow
circuit type)
Rise Time
Ramp Length
All other alarms
B. Options Menu
Modify the settings in the Options menu to match those shown
below in Table 2.
Setting
Menu Access
Detailed View
All other settings
C. Turn Device Power On
Press the Start/Stop button on the front of the ventilator. The
system will begin operating using the defined ventilation
settings.
Value
S/T
Off
20 cmH
O
2
4 cmH
O
2
45%
12 BPM
1.6 seconds
Auto-Trak
6.0 l/min
On
20%
1
Off
Off
Value
Full
On
Discretionary
Table 2 - Ventilator Settings
in the Options Menu
Chapter 10 System Checkout Procedures
139