PHILIPS Respironics Trilogy 202 User Manual page 83

Ventilator
Table of Contents

Advertisement

74
Indicator
Leak
Measured Settings Panel
All of the indicators that may appear on the Measured Settings panel
(available only in Detailed view), are described in the following table.
Indicator
PIP
I:E Ratio
Peak Flow
MAP
FiO
2
MinVent
Trilogy 202 clinical manual
This indicator displays the total leak (non-returned flow), between
the unit outlet and the patient, averaged over the previous
breath. This only appears in detailed view when the Passive
Circuit is selected.
Peak Inspiratory Pressure displays the maximum pressure
delivered to the patient during the previous breath.
Displays a comparison of the time spent in inspiration to the time
spent in expiration during the previous breath.
Displays the maximum inspiratory flow delivered to the patient
during the previous breath in l/min BTPS.
Displays the Mean Airway Pressure, which is the weighted
average of pressure in the patient's airway over 6 breaths.
Displays the FiO
setpoint.
2
Minute ventilation displays the amount of air delivered to the
patient over the last minute in l/min BTPS.
Description
Measured Settings Panel
Description