146
C. Modify Ventilator Alarm Settings
Modify the Low Inspiratory Pressure setting to match the one
shown below in Table 14.
Setting
Low Inspiratory Pressure
D. Verify Reset
Wait 40 seconds and verify the following auto-reset conditions:
The High Priority audible indicator has stopped
sounding
The red light on the
button has stopped flashing
7. Verify the O
2
This procedure verifies that the Low Oxygen Flow alarm and the Low
Oxygen Inlet Pressure alarm are working properly. These alarms
apply to all circuit types. It assumes that you have attached the
test lung, verify the ventilator settings, and turned on the ventilator
power as described in the Initial Setup section.
Setup
A. Connect an external O
recommended manufacture's guidelines) in-line with the patient
tube. Make sure the O
proceeding.
B. Connect the ventilator's oxygen input port to a source of high
pressure O
(50 psi nominal). Turn on O
2
Ventilator Settings
Set the ventilator FiO
Trilogy 202 clinical manual
6 cmH
Alarm Indicator/Audio Pause
Blending Operation
Monitor (in accordance with the
2
monitor is properly calibrated before
2
setting to 45%.
2
Value
O
2
flow to the ventilator.
2
Table 14 - Modify Low
Inspiratory Pressure Alarm
Setting