Visual Inspection; Initial Setup; Settings And Alarms Tests - PHILIPS Respironics Trilogy 202 User Manual

Ventilator
Table of Contents

Advertisement

138

Visual Inspection

1. Verify that the enclosure is not broken and that all applicable
screws are in place.
2. Verify that the device handle, SD Card door, and detachable
battery are secure and in good working order.
3. Verify that the rubber feet are on the bottom of the device.

Initial Setup

1. Connect the power cord to the device and then to an AC outlet.
2. Attach the test lung to the patient connection end of the desired
circuit (Active PAP, Active Flow, or Passive).
3. Connect the device to a suitable high pressure O
4. Connect and set up an external O
manufacturer's instruction manual.
5. Follow the instructions in Chapter 5 to access the Setup Screen.

Settings and Alarms Tests

Complete the following steps to set up the settings and alarms tests.
1. Setup
A. Settings And Alarms Menu
Modify the settings in the Settings and Alarms menu to match
those shown below in Table 1. If necessary, refer to Chapter 5 for
instructions on modifying ventilator settings.
Setting
Dual Prescription
Circuit Type
Trilogy 202 clinical manual
source.
2
monitor per the
2
Value
Off
Active PAP, Active Flow or
Passive
Table 1 - Ventilator Settings
in the Settings and Alarms
Menu