Viewing The Alarm Log - PHILIPS Respironics Trilogy 202 User Manual

Ventilator
Table of Contents

Advertisement

Minute
– The minute defaults to the current minute. The
adjustable range is from 0 – 59.
IP Address Mode – You can change the IP address
mode to either DHCP or Static, depending on the type of
network you are using (if applicable).
Operational Hours – The operational hours displays the
total number of hours that the device blower has been
on since the last time this value was reset. You can reset
this value to zero (0) if desired. The Operational Hours
shown here differs from the Blower Hours shown on the
Information screens. The Blower Hours displayed in the
Information screen is the total number of hours that the
blower has been working over the life of the device. You
cannot reset this value.

Viewing the Alarm Log

From the Main Menu screen, you can select Alarm Log to access the
Alarm Log screen. An example is shown below.
The alarm log displays the alarms in chronological order with the
most recent events displayed first. It lists the 20 most recent alarms
or messages that appeared on the device display. When the device is
in Limited Menu access mode, the alarm log cannot be cleared. It can
be cleared when in Full Menu access mode. Depending on how many
alarms have occurred, the alarm log may be several pages long. The
entries in the alarm log use the same names that displayed when the
alarm initially occurred and was displayed in the Alarm View.
Note: In the Alarm Log screen,
the 1/2 shown in the Menu
banner indicates that page 1
of 2 alarm log pages is being
viewed at this time.
Alarm Log Screen
Note: In Full Menu access mode,
you can press the Right (Clear)
button to clear the alarm log if
desired.
Chapter 5 Viewing and Changing Settings
97