60

Battery Disposal

Dispose of the batteries in accordance with local regulations.

First Time Use

When setting up the device for the first time or after annual service,
apply AC power to the ventilator before turning on the blower.
Attempting to use the ventilator without first applying AC power,
such as installing a detachable battery pack and starting the blower,
will cause the internal battery to be displayed in red as an empty
battery. When in this state, the internal battery will not be usable
until AC power is applied.

Connect the Breathing Circuit to the Ventilator

Complete the following steps to set up your patient circuit.
1. If you are using a viral/bacteria filter (highly recommended), connect
one end of the flexible tubing to the outlet of the bacteria filter, and
connect the bacteria filter inlet to the breathing circuit connection
located on the side of the ventilator.
If you are not using a viral/bacteria filter (not recommended), connect
the tubing directly to the device's breathing circuit connection.
Flexible Tubing
Bacteria Filter
2. Connect the other end of the flexible tubing to a separate
exhalation device.
a. If your circuit type is Passive and you are using a Passive
Exhalation Device:
1. Connect the flexible tubing to the 22mm O.D. end of the
passive exhalation device.
Trilogy 202 user manual
Trilogy Breathing Circuit
Connection
WARNING
To prevent patient or ventilator
contamination, we recommend
you use a Respironics-approved
main flow bacteria filter (Part
Number 342077) on the patient
gas outlet port. Filters not
approved by Respironics may
degrade system performance.
Connecting the Breathing
Circuit to the Device
Note: The device can be used
with reusable or disposable
circuits. For detailed instructions
on how to set up your device
using a disposable circuit, refer
to the instructions included with
the disposable circuit.