PHILIPS Respironics Trilogy 202 User Manual page 131

Ventilator
Table of Contents

Advertisement

122
Alarm
Priority
Check External
Info
Battery
Battery
Info
Depleted
External Battery
Info
Disconnected
Detachable
Info
Battery
Disconnected
Start On Battery Info
Card Error
Info
Trilogy 202 clinical manual
Audible
Visual (Alarm
Indicator Button
and Display)
"Check External
Battery" message
"External Battery
Depleted" or
"Detachable Battery
Depleted" message
appears
"External Batt
Disconnected"
message, and a box
appears around
battery in use.
"Detachable Batt
Disconnected"
message, and a box
appears around
battery in use.
"Start On Battery"
message appears.
A "Card Error"
message appears.
Device
Clinician Action
Action
Continues to
If the alarm continues after
the measures above have
operate
been taken, place patient
on alternate source of
ventilation and have device
serviced.
Continues to
Replace depleted battery
with another or switch to AC
operate
power.
Switches to
Check external battery and
recharge if necessary. Restore
alternate
AC power if available.
power source
Switches to
Check detachable battery
and recharge if necessary.
alternate
Restore AC power if available.
power source
Continues to
Restore AC power when
available. If alarm continues,
operate
replace AC power cord
to determine if there is a
problem with the cord.
Continues to
Remove SD Card and use
another card, if available.
operate
Ensure card meets
specifications. Check write-
protect switch on card. If
condition persists, contact an
authorized representative of
Philips Respironics.