62
B. Attach the proximal pressure line (.476 cm or 3/16"
line) to the proximal pressure port as shown in the
next illustration.
C. Connect the other end of the proximal pressure line
to the Trilogy Universal Active PAP Tube Adapter or
equivalent as shown in the next illustration.
D. Connect the exhalation valve line (.317 cm or 1/8" line)
to the exhalation valve port on the top of the active
exhalation device with proximal pressure.
E. Connect the other end of the exhalation valve line
to the exhalation valve port on the Universal Porting
Block.
Exhalation Valve Port
Proximal
Pressure
Port
Refer to the instructions included with the Active Exhalation
Device for more information.
c.
If your circuit type is Active Flow:
1. Connect the Flow Sensor to an active exhalation device as
shown below.
Flow Sensor
Trilogy 202 user manual
Proximal
Pressure
Trilogy Universal Active
Line
PAP Tube Adapter
Exhalation Valve Line
Exhalation Valve Port
on Universal Porting
Block
Universal Porting Block
Note: Make sure that
components marked with an
arrow are oriented properly.
When attaching an active
exhalation device with proximal
pressure to the patient, make
sure the proximal pressure port
faces away from the patient.
Note: The
symbol
appears next to the Exhalation
Valve port on the Universal
Porting Block to indicate where
the active exhalation valve line
connects.
Connecting an Active Device
with Proximal Pressure
Note: Only use the Respironics-
approved Flow Sensor with this
device.
Connecting the Flow Sensor