PHILIPS Respironics Trilogy 202 User Manual page 111

Ventilator
Table of Contents

Advertisement

102
3. Select Cancel to cancel the prescription update process and
return the screen to the initial state before the prescription
update started. Select Page to review the entire prescription.
The Menu Banner will reflect the prescription being updated.
One of the following screens displays once the entire
prescription has been reviewed:
4. Select Cancel to cancel the prescription update process and
return the screen to the initial state before the prescription
update started. Select OK to complete the prescription update
and display the Prescription Change confirmation screen.
If the SD card is removed at any time during the prescription
update, the process aborts and the screen returns to the initial
state before the prescription update started.
A confirmation screen will appear if errors occur during this process.
The following table summarizes the prescription errors, their possible
causes, and the actions to take.
Trilogy 202 clinical manual
Note: Make sure you closely
review the prescription and
confirm that all settings are
correct.
Ventilator Off Prescription
Confirmation Screen
Ventilator On Prescription
Confirmation Screen
Note: If both prescriptions of
a dual prescription are being
updated, you must accept the
primary prescription using
the OK button before being
able to review the secondary
prescription. Both prescriptions
have to be accepted before any
changes are made.

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents