5. Press the Offset Key. The Digital
Readout will display the Tip Offset
Constant value (which is "0" as received
from the factory). If the Offset Key is
immediately pressed again, or if no other
operation occurs within 5 seconds, the
Digital Readout will revert to the
Temperature Display Mode (normal
operation).
If a Password has been previously programmed
into the system, a message will be displayed on
the Digital Readout (when a setting change is
attempted) requiring the operator to enter the
correct Password before adjusting the system
parameters.
Refer to "Password" in the
"Operation" portion of this manual.
6. If the Digital Readout reverts to the
Temperature Display Mode (normal
operation), press the Offset Key once to
enter Tip Offset Mode. Immediately press
and hold the Scroll Up ( ) Key. Observe
the displayed Tip Offset Constant value
increase, first in 1° and then in 10°
increments. Release the key when the Tip
Temperature Offset value reads "033" for
°C ("060" for °F).
7. While still in the Tip Temperature Offset
Mode, press and hold the Scroll Down
( ) Key. Observe the displayed Tip
Offset Constant decrease first in 1° and
then in 10° increments. Release the key
when the Digital Readout displays "028"
for °C ("050" for °F).
NOTE
Operation
XXX
XXX °C
033
C
XXX
028
C
XXX