Installation Checkout - GE Cardiolab II Plus Service Manual

Amplifier
Table of Contents

Advertisement

Operation

Installation Checkout

3-4
Once the appropriate cable connections have been made to the
CardioLab II Plus Amplifier, the communication indicators on the front
of the amplifier should turn on. If the indicators do not turn on, check the
fiber optic cable for cracks and wear.
If the XMT indicator is on, the CardioLab II Plus Amplifier is
transmitting data to the acquisition computer. The XMT indicator will
turn on regardless if the transmission of data is successful or not. The
RCV ERR indicator will turn on if the amplifier is having difficulty
transmitting data to the computer.
The RCV CMD indicator will turn on when the acquisition computer is
sending data to the CardioLab II Plus Amplifier. If the RCV ERR
indicator is on, the acquisition computer may be having difficulty
transmitting data to the computer.
The RCV ERR indicator may also turn on if the fiber optic cable has been
connected to the amplifier incorrectly.
Refer to the manuals listed in the table in the Related Manuals section in
Chapter 1 for information on configuring and diagnosing the CardioLab
II Plus Amplifier using the acquisition computer.
Cardiolab® II Plus Amplifier Manual
2002938-002E