1.2 Combination possibilities
This prosthetic component is compatible with Ottobock's system of
modular connectors. Functionality with components of other manufac­
turers that have compatible modular connectors has not been tested.
Limited combination options for Ottobock components
This product can be combined with the following Ottobock compon­
ents:
Socket attach­
5R1=1, 5R1=2, 5R1=6, 5R2, 5R6=1, 5R6=2,
ment block
5R6=3
Socket adapter
4R22, 4R37, 4R51, 4R55, 4R95
2 Intended use
2.1 Indications for use
The product is to be used solely for lower limb prosthetic fittings.
2.2 Area of application
The product must be used in TF prostheses only.
Approved for a body weight of up to 125 kg.
2.3 Environmental conditions
Storage and transport
Temperature range –20 °C to +60 °C (–4 °F to +140 °F), relative
humidity 20 % to 90 %, no mechanical vibrations or impacts
Allowable environmental conditions
Temperature range: –10 °C to +45 °C
Moisture: relative humidity: 20% to 90%, non-condensing
Unacceptable environmental conditions
Chemicals/liquids: fresh water, salt water, perspiration, urine,
acids, soapsuds, chlorine water
Solids: dust, sand, highly hygroscopic particles (e. g. talcum)
8
2.4 Service life
In general, all modular adapters are tested by the manufacturer for
3 million load cycles. Depending on the user's activity level, this cor­
responds to a maximum service life of 5 years.
3 Safety
3.1 Explanation of warning symbols
Warning regarding possible risks of accident or
CAUTION
injury.
Warning regarding possible technical damage.
NOTICE
3.2 General safety instructions
CAUTION
Excessive strain on the product
Risk of injury due to breakage of load-bearing components
► Use the product according to the specified area of application
(see Page 8).
CAUTION
Unallowable combination of prosthetic components
Risk of injury due to breakage or deformation of the product
► Only combine the product with prosthetic components that are
approved for that purpose.
► Based on the instructions for use of the prosthetic components,
verify that they may be combined with each other.
CAUTION
Use under unallowable environmental conditions
Risk of injury due to damage to the product
► Do not expose the product to unallowable environmental condi­
tions.

Advertisement