5.2 Installation in the modular prosthesis
CAUTION
Improper assembly of the screw connections
Risk of injury due to breakage or loosening of the screw connec­
tions
► Clean the threads before every installation.
► Apply the specified torque values.
► Follow the instructions regarding the length of the screws and
about how to secure the screws.
Mounting on socket attachment block
Recommended tools:
>
710D20 torque wrench, 636K13 Loctite®
1) Select the correct cap screws for the socket attachment block or
the lamination disc (see selection table).
2) Position the adapter plate with the slotted holes on the socket
attachment block or the lamination disc.
3) Adjust the desired posterior displacement.
4) Apply Loctite® to the cap screws, slide them through the slotted
holes and tighten them crosswise (torque value 10 Nm).
Socket attachment block,
lamination disk
5R2
5R1=1, 5R1=2, 5R1=6
5R6=1, 5R6=2, 5R6=3
Fastening the socket adapter
► To connect the adapter plate to a socket adapter, use
501S41=M6x16 countersunk screws.
Required materials: 710D20 torque wrench, 636K13 Loctite 241
>
1) If the pyramid receiver or the pyramid adapter is rotatable:
Place the pressure plate on the adapter.
2) Place the adapter on the connecting component.
10
Cap screw
501T61=M6x12
501T61=M6x25
501T61=M6x30
3) Optional: Align the pyramid receiver or the pyramid adapter.
4) Secure the screws with Loctite.
5) Screw in the 2 posterior countersunk head screws and tighten
them (tightening torque: 12 Nm).
6) Screw in the 2 anterior countersunk head screws and tighten
them (tightening torque: 12 Nm).
6 Cleaning
1) Clean the product with a damp, soft cloth.
2) Dry the product with a soft cloth.
3) Allow to air dry in order to remove residual moisture.
7 Maintenance
► A visual inspection and functional test of the prosthetic compon­
ents should be performed after the first 30 days of use.
► Inspect the entire prosthesis for wear during normal consulta­
tions.
► Conduct annual safety inspections.
8 Disposal
This product may not be disposed of with regular domestic waste in
all jurisdictions. Disposal that is not in accordance with the regula­
tions of the country where the product is used may have a detrimental
impact on health and the environment. Please observe the information
provided by the responsible authorities in the country of use regarding
return, collection and disposal procedures.
9 Legal information
All legal conditions are subject to the respective national laws of the
country of use and may vary accordingly.
9.1 Liability
The manufacturer will only assume liability if the product is used in
accordance with the descriptions and instructions provided in this
document. The manufacturer will not assume liability for damage

Advertisement

Table of Contents
loading

Table of Contents