Table Of Functions - KYOWA WGA-680A Series Instruction Manual

Instrumentation amplifier
Hide thumbs Also See for WGA-680A Series:
Table of Contents

Advertisement

4-12-3. Table of functions

No.
Functions
Initialize
F-01
Original
value
F-02
Actual load
calibration
F-03
Sensitivity
registering
calibration
Parameters
Digits
Rated value
-99999 to +99999
Decimal point
10e0
position
10e1
10e2
10e3
10e4
Rated output
-3.2000 to +3.2000
Rated value
-99999 to +99999
Decimal point
10e0
position
10e1
10e2
10e3
10e4
Initial
Options/ranges
values
4000
None
None
The 10th digit
lights up.
The 100th digit
lights up.
The 1000th digit
lights up.
The 10000th
digit lights up.
2.0000
4000
None
None
The 10th digit
lights up.
The 100th digit
lights up.
The 1000th digit
lights up.
The 10000th
digit lights up.
48
IM-A-1048H
Remarks
The current input value (mV/V) is
displayed.
Linked to the sensitivity registering
calibration.
Decimal point: None
The decimal point is displayed at
the 10th digit.
The decimal point is displayed at
the 100th digit.
The decimal point is displayed at
the 1000th digit.
The decimal point is displayed at
the 10000th digit.
Rated input (mV/V)
The indicated value when the input
is the rated input.
This parameter is linked to the rated
value of the actual load calibration.
Decimal point: None
The decimal point is displayed at
the 10th digit.
The decimal point is displayed at
the 100th digit.
The decimal point is displayed at
the 1000th digit.
The decimal point is displayed at
the 10000th digit.