FUNCTION
SELECTING
MODE
F-08
Additional value Set the additional value and
press the SET key.
The sum of the measured value and pre-set value is
displayed.
To display the pre-set value on the indicator, execute the
DIGITAL ZERO function with the additional value being
set.
F-09
Automatic
(See below figures)
ZERO
compensation
(1) ZERO compensation range
(2) Determination time
(3) Zero-near-zero
F-10
Hysteresis
Set the value and press the SET key.
For details, see separate volume of
F-11
BCD output
"INSTRUCTION MANUAL (FOR BCD/RS-232C/
Logic
RS-485/TEDS OPTIONS) ".
Key operation and WGA-680A movement
in SETTING mode
Range: ±99999 (Initial value: 0)
Set the value and press
the SET key.
Range: 0 to 99999 (Initial value: 0)
Set the time and press
the SET key.
Range: 0.0 to 9.9 (sec.) (Initial value: 0.0)
Set the value and press
the SET key.
Range: 0 to 9 (Initial value: 0)
Range: 0 to 99999 (Initial value: 0)
43
IM-A-1048H
Functional description,
supplements
A function that adds the
desired numeric value to
the measured value for
displaying.
Zero tracking
When the
determination time is
passed and the
absolute value of the
measured value is
within the zero
compensation range,
this function executes
the DIGITAL ZERO
function and
automatically makes
the indicated value
"0".
Zero-near-zero
When the measured
value is within the
zero-near-zero range,
this function
automatically makes
the indicated value
"0".
A function that sets the
common hysteresis to four
comparators.
BCD/TEDS model only.