KYOWA WGA-680A Series Instruction Manual page 37

Instrumentation amplifier
Hide thumbs Also See for WGA-680A Series:
Table of Contents

Advertisement

(2) Hold display mode
Key operation
When the HOLD key is pressed, the WGA-680A holds the value at that time. The detecting interval starts
after the lapse of delay time.
When the HOLD key is pressed for the second time, the holding interval ends. The WGA-680A displays
values as normal.
Control input
While the HOLD command is ON, the WGA-680A holds the value, detected when turning ON the HOLD
command. The detecting interval starts after the lapse of delay time.
When the RESET command is turned ON, the holding interval ends. The WGA-680A displays values as
normal.
NOTE
When selecting the "Hold comparison mode", the WGA-680A compares values
during the holding interval only.
36
IM-A-1048H

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents