Table of Contents

Advertisement

4-6. PATTERN FUNCTION
You are able to register the [HH]/[HI]/[LO]/[LL] limit values in 4 operation patterns (pattern 0 to pattern 3). You
are able to select the target pattern by the key operations.
a) To select operation patterns
Select the operation patterns.
For details, see "F-20" of "4-11-1. Various functions in FUNCTION SELECTING".
b) To write the pre-set values
Set the comparator values after changing the operation patterns (pattern 0 to pattern 3).
The comparator pattern that you set is registered the operation patterns.
NOTE
The WGA-680A keeps the value after turning OFF the power.
4-7. LEVEL TEST FUNCTION
The LEVEL TEST function displays the pre-set values as the indicated value. The LEVEL TEST function is used
for adjusting the external connected devices or testing the output of the WGA-680A.
a) To set
Display additional function
0 (Disabled)
1 (Enable)
LEVEL TEST values
LEVEL TEST value
Initial value
For details, see "F-22" of "4-11-1. Various functions in FUNCTION SELECTING".
b) To operate
The pre-set value is displayed while the LEVEL TEST ON/OFF is "1: LEVEL TEST ON".
For details, see "F-22" of "4-11-1. Various functions in FUNCTION SELECTING".
: Displays the pre-set value regardless of the measured value.
: Displays the measured value added to the pre-set value.
: ±99999
: 1000
26
IM-A-1048H