KYOWA WGA-680A Series Instruction Manual page 26

Instrumentation amplifier
Hide thumbs Also See for WGA-680A Series:
Table of Contents

Advertisement

4-5-3. To set the [LL] limit value and [LO] limit value
1) Press the LO LL key for 2 seconds in the MEASURING mode. The current value of the [LO] comparator is
displayed and the highest-order digit flickers. At this time, the [LO] LED flickers.
3) To set the [LO] limit value, press the SET key.
4) The current value of the [LL] comparator is displayed and the highest-order digit flickers. At this time, the
[LL] LED flickers.
6) To set the [LL] limit value, press the SET key.
*Initial values of the [LL] limit value and [LO] limit value are -99999
4-5-4. To set the hysteresis
You are able to set the hysteresis width that turns OFF comparators which was once turned ON.
The set hysteresis is common in four comparators.
For details, see "F-10" of "4-11-1. Various functions in FUNCTION SELECTING".
4-5-5. To assign comparators
You are able to assign "High limit" or "Low limit" to each comparator.
Suppose the [LO] comparator is assigned "High limit" and [LL] comparator is assigned "High limit" too. You
are able to set them as the level comparator.
For details, see from "F-23" to "F-26" of "4-11-1. Various functions in FUNCTION SELECTING".
25
IM-A-1048H