(2) Hold display mode
Key operation
When the HOLD key is pressed, the WGA-680A holds the value at that time. The detecting interval starts
after the lapse of delay time.
When the HOLD key is pressed for the second time, the detecting interval ends and the holding interval
starts. The WGA-680A displays the detected value.
When the HOLD key is pressed for the third time, the holding interval ends. The WGA-680A displays
values as normal.
Control input
While the HOLD command is ON, the WGA-680A holds the value, detected when turning ON the HOLD
command. The detecting interval starts after the lapse of delay time.
When the HOLD command is turned OFF, the detecting interval ends and the holding interval starts. The
WGA-680A displays the detected value.
When the RESET command is turned ON, the holding interval ends. The WGA-680A displays values as
normal.
NOTE
When selecting the "Hold comparison mode", the WGA-680A compares values during the holding
interval only.
34
IM-A-1048H