FUNCTION
SELECTING
MODE
F-06
Analog output
(See below figures)
(D/A) scaling
(1) Automatic scaling
(2) ZERO indicated value
(3) FULL-SCALE indicated value
F-07
Displays the following items (1). (2). (3).
Smoothing
(See below figure)
(1)Number of moving average times (2) Minimum scale
(3) Analog filter
Select and change values in items (1), (2), and (3)
And press the SET key.
NOTE
When the number of moving average times is set, response
of the analog output is deteriorated.
Key operation and WGA-680A movement
in SETTING mode
0: Manual setting
1: Automatic setting (initial value)
Select the value and press the SET key.
Set the indicated value to have
the output "0 V" or "4 mA" and
press the SET key.
Range: ±99999 (Initial value: 0)
Set the indicated value to have
the output "10V" or "20 mA" and
press the SET key.
Range: ±99999 (Initial value: 4000)
0 0 time (Initial value)
1 2 times
2 4 times
3 8 times
4 16 times
5 32 times
6 64 times
7 128 times
8 256 times
9 512 times
A 1024 times
b 2048 times
0 300 Hz (Initial value)
1 100 Hz
2 30 Hz
3 10 Hz
0 1 (Initial value)
1 2
2 5
3 10
4 20
5 50
6 100
7 200
8 500
9 1000
(1) (2) (3)
42
IM-A-1048H
Functional description,
supplements
With the "1: Automatic
setting", the calibrated
value is automatically set
as the FULL-SCALE
indicated value.
For details, see "4-11.
ANALOG OUTPUT
(D/A) FUNCTION".
A function that stables the
indication by smoothing
the varying signals from
the transducer.
(1) Number of moving
average times
The number of moving
average times that
averages the measured
values for displaying.
(2) Minimum scale
Minimum unit that
varies the indicated
value.
(3) Analog filter
A function that cuts the
higher frequencies
generated from
transducers.