To Connect [Analog Output] Terminals - KYOWA WGA-680A Series Instruction Manual

Instrumentation amplifier
Hide thumbs Also See for WGA-680A Series:
Table of Contents

Advertisement

3-6. TO CONNECT [ANALOG OUTPUT] TERMINALS

The pins and signals of the [ANALOG OUTPUT] terminals are as follows.
COM
: Common
Vo
: Voltage output terminal (±10 V)
Io
: Current output terminal (4 to 20 mA)
NOTE
Countermeasures against noise for [INPUT]/[OUTPUT]/[ANALOG OUTPUT] terminals
For more than 1 m long wiring, use twisted cable.
For more than 3 m long wiring, use a shield cable and connect the shield cable to the GND terminal.
For more than 30 m long wiring or for using the WGA-680A outdoors, be sure to take additional
countermeasures against noise.
For connecting multiple wires, be sure to twist them together in advance. After twisting, make sure
the wire is satisfied with the diameter to connect the terminal block.