27 Help and error messages
Help messages
HELP
MEANING
CODE
H 90.2
ActiveAqua prefilter expired.
H 90.3
ActiveAqua water container is empty.
H 90.4
ActiveAqua water container overflow.
H 90.5
ActiveAqua annual maintenance overdue.
H 90.6
Measured chlorine level low.
H 90.7
Measured chlorine level OK.
H 90.8
Measured chlorine level high.
H 91
Stop (Help) button pressed.
H 92
Help (Stop) button pressed.
H 93
This button has no function at the moment
because there is no motorised light.
H 94
Clock calibration failed. Maximum
calibration range exceeded.
H 95
Patient chair automatic position cannot be
reached because legrest is unlocked or
armrests are blocking movement.
H 96
PlanID tag already in use.
H 96.1
Different user already logged in.
H 97
Failed to assign PlanID tag to user.
H 98
User not found.
H 98.1
User login failed.
H 98.2
RFID unknown.
H 99
Unit has been switched off without
performing long flushing after water has
been used. Perform long flushing.
HE200 Detected new software update. Confirm to
install?
HE201 Detected new software update SD-card.
Confirm to install 5.3.0.9.R ?
148
Planmeca Compact i
ACTION OR EXPLANATION
Switch off the dental unit. Replace the
prefilter. After switching on the dental unit
press the Reset prefilter button to confirm
the prefilter replacement.
To temporarily ignore this notification, select
any instrument or press Cancel.
Water container fill is slow or the low level
sensor does not sense water.
Check ActiveAqua level sensors and water
container.
Annual ActiveAqua maintenance must be
performed. Contact service.
Normal use chlorine level increased by
+15%.
No change to normal use chlorine level.
Normal use chlorine level decreased by
-5%.
PlanID tag assigned to another user.
The unit has rejected a login request
because a different user is already logged
in.
Check Romexis connection.
User not found with this PlanID tag. Create
a new user or select an existing Romexis
user and assign the PlanID tag.
Retry login or reboot the unit. If problem
persists, contact service.
The RFID is unknown to the system.
This help message can be disabled, contact
your Planmeca dealer.