27 Help and error messages

27.3.8 Operating light related error messages

27.3.9 Separator system related error messages

27.3.10 Patient chair related error messages

158
Planmeca Compact i
Foot control related error messages
ERROR
EXPLANATION
CODE
E54.9
Foot control calibration error.
E55
Wireless foot control connection lost.
Operating light related error messages
ERROR
EXPLANATION
CODE
E56
Operating light erroneously connected or short circuit in push
button.
E57
Incompatible SingLED software.
E58
Operating light bulb burned out.
Replace bulb.
Separator system related error messages
ERROR
EXPLANATION
CODE
E59.1
Separator reports functional error.
Refer to Metasys separator documentation.
E59.2
Separator reports functional error.
Refer to Dürr separator documentation.
E60
Separator is not responding or separator cable is faulty.
E61
Separator is flooded with too much water.
E62
Newly attached separator is of new type and cannot be
controlled by MCB.
E63.1
Water management system (WMS) seems to be installed,
although it should not be (according to service mode).
E63.2
No Water management system (WMS) is installed, although it
should be according to service mode.
Patient chair related error messages
ERROR
EXPLANATION
CODE
E64
Lift motor position potentiometer or its cable is faulty or
disconnected.
E65
Lift motor does not run or no signal from position
potentiometer.