24 Cleaning and disinfection
24.5.2 After each patient
Steps
Results
24.5.3 After each working day
Steps
124
Planmeca Compact i
1. Remove the used aspirating tips.
2. For hygienic and operational reasons, empty one glass of water (100 -
200 ml) with each suction tube by aspirating water and air through the
suction handpiece. Do this even if only the saliva suction tube has been
used.
3. Wipe the suction handpieces with Planmeca approved surface
disinfectant.
4. Wipe the suction holder / Flexy-holder, including the tablet holder, with
Planmeca approved surface disinfectant.
NOTE
Do not use a spray disinfection solution for the suction arms and holders.
NOTE
Clean the tablet according to the instructions given by the manufacturer.
1. Disinfect the suction system either by running the suction cleaning
program or by using OroCup. For information on the suction cleaning
program, see section "Suction cleaning" on page 115.
2. When the suction system has been rinsed, disinfect the suction
handpieces in a washer-disinfector at 93°C, then optionally autoclave
them at 134°C.
3. Flexy-holder: Disinfect the supplementary holders, suction tube holders
(including rollers) and instrument holder in a washer-disinfector at 93°C.
4. Flexy-holder: Wipe the holder for the intraoral scanner with Planmeca
approved surface disinfectant.