27 Help and error messages

27.1 Overview

27.2 Help messages in short form

Help messages
HELP
MEANING
CODE
H 1
Amalgam collector is 95% full.
H 2
Amalgam collector is 100% full.
H 3
Patient chair cannot be driven down
because safety switch is activated.
H 4
This left-hand instrument position allows
only syringe.
H 5
This instrument position does not accept
syringe.
H 6
Patient chair cannot be driven when
instrument is in use.
H 7
Patient chair up/down movement is not
allowed when cuspidor door is open.
User's manual
The dental unit displays two types of safety messages: help messages and
error messages.
Help messages provide helpful information for the user. They are displayed,
for example, if you are using the unit or instrument incorrectly, or if the
function is not allowed for some reason. The help message disappears
automatically when the situation is corrected.
Error messages warn of a fault in the dental unit and require actions from the
user. To correct the error situation, and thus to close the error message,
follow the instructions in the message.
NOTE
The WCS related help messages can be disabled. Contact your Planmeca
dealer.
27 Help and error messages
ACTION OR EXPLANATION
Replace or empty the collector as soon as
possible, or at least before the next patient.
Pressing the program button allows to
continue the operation until the collector is
100% (totally) full.
Replace or empty the collector immediately.
The suction system cannot be used before
the collector is emptied or replaced.
Check that the space under the seat is free
and nothing is touching the plate at the
bottom. The chair can be driven normally
after the possible obstruction has been
removed.
The leftmost position in the console is
reserved for the syringe only. Replace this
instrument with a syringe. Place the other
instrument to one of the four right-hand
places.
Remove the syringe from this position and
replace it with any other kind of instrument.
The syringe must be placed in the leftmost
place.
Stop the instrument (release the foot control
pedal) to be able to run the chair.
Close the cuspidor door to be able to run
the chair up/down or to a pre-programmed
position.
Planmeca Compact i