-->Continue from the previous page
Function
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25
0 0 0 0 CR LF
2
N CR LF
N CR LF
R CR LF
・Reset the SPIN value and SPAN setting value for clearing Error 5.
5 0 0 0 CR LF
1
N CR LF
2 0 0 0 CR LF
1
N CR LF
R CR LF
Transfer data
・Reset the SPAN set value.Respond the setting value.
・For details of Error 5, refer to "7-1".
 Perform N when no problem.
 Reset the SPN setting value when problem happened.
・Read the SPIN value and SPAN setting value when N is
performed, and return to the beginning.
・Return to normal screen after performing R and SAV.
・Reset SPIN value when Error 6. Respond the setting
value.
・Read the SPIN value when N is performed. Respond the
SPAN setting value after changing
・Reset the SPAN setting value. Respond the setting value.
・Read the SPIN value and SPAN setting value when N is
performed, and return to the beginning. Error 5 is cleared.
Setting again is possible.
・Return to normal screen after performing R and SAV.
Chara
cter
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25
2 0 0 . 0 0 CR LF
S P A N
5 CR LF
E r r o r
1 . 5 0 0 0 m V / V CR LF
S P I N
0 . 0 0 0 5 m V / V CR LF
Z E R O
Y E S
5 CR LF
E r r o r
1 . 5 0 0 0 m V / V CR LF
S P I N
1 . 5 0 0 0 m V / V CR LF
S P I N
2 0 0 . 0 0 CR LF
S P A N
1 2 0 . 0 0 CR LF
S P A N
0 . 0 0 0 5 m V / V CR LF
Z E R O
92/118