b) SPAN set value
・Set the indicated value when the SPIN value (rated-output value) is
acquired and finalized.
・Acquire set value and finalize with "SET" key.
(Set value becomes blue simultaneously with finalizing.)
SPAN set value
・The acquired SPAN set value is checked.
Resolution in this Meter is 10000 at the time of a 1.000mV/V value.
k1= SPAN set-value / (SPIN value × 10000)
・Error 5 is displayed at the time of k1 > 1.
・When no problem occurs at Error 5, press "SET" key to acquire the set
value, which can be directly used.
When a problem occurs, press "C" to set the set value to "0", or press
"▲" to return to the set-value select screen of the above-mentioned 6)
and then press "SPAN" key for resetting.
If Error 5 still occurs, it is necessary to press the "▲" key twice and
to perform resetting from 5) Zero Calibration.
7) Actual Load
・The measured value (SPAN) when actual load is applied to the strain
gauge sensor and the SPAN set value (indicated value) at the time are
acquired.
・Acquisition range of measured value (SPAN value) -3.0 to 4.0mV/V.
・Setting range of SPAN set value (indicated value) 100 to 99999.
・Set it on the measured-value acquisition + 10-key input screen.
* Caution
When「6. Basic Function Setting (Setup) 4) Motion Detect」is active,
measured value becomes red while analog output is not stable.
Setting is possible even during unstable state.
・Acquire both SPAN set value and measured value (SPAN value) and
finalize with "SET" key.
(Set value becomes blue simultaneously with finalizing.)
SPAN set value
In case of SPAN set value error
When SPIN value is error