Itemized setting screen 4/5
・A value obtained by subtracting ZERO value from the acquired
measured value (SPAN value) is converted into SPIN value (mV/V
value) and it is checked. In the following cases, it is judged an error and
resetting is performed.
Error 3:When it is smaller than 0.1mV/V value
Error 4:When it is larger than 3.0mV/V value
・Perform resetting by pressing the "▲" key and then pressing the
"Actual Load" key on the above-mentioned itemized setting screen.
Check the condition with actual load applied and then press the "SET"
key.
・When SPAN set value is error 5
・Resolution in this Meter is 10000 at the time of a 1.0000mV/V value.
k1= SPAN set value / (SPIN value ×10000)
・Error 5 is displayed at the time of k1 > 1.
・When no problem occurs at Error 5, press "SET" key to acquire the set
value, which can be directly used.
When a problem occurs, press "C" to set the set value to "0", or press
"▲"and press "Actual Load" key on the above-mentioned itemized
setting screen for resetting.
If Error 5 still occurs, it is necessary to press the "▲" key twice and to
perform resetting from 5) Zero Calibration.
8) Decimal Point
・Set the decimal point position displayed on the set-value select screen.
・Setting items : 0, 0.0, 0.00, 0.000, 0.0000
9) Digital Shift
・Function provided to simply shift the indicated value.
Indicated value = (Internal indicated value) + (±digital shift value)
・Set it on the 10-key input screen.
・Setting range : ±99999
10) Unit
・Set the unit corresponding to the indicated value.
・Set it on the set-value select screen.
・79 different setting items (including those without display)
Refer to「29. Setting of Units.」
11) Moving Average
・This function is provided for the moving average of data after A/D
conversion so as to reduce the fluctuation of the indicated values. As
the number of times of moving average is increased, the indicated
value is stabilized. However response becomes slow.
・Set it on the set-value select screen.
・Setting items : OFF, 2, 4, 8, 16, 32, 64, 128, 256, 512, 1024 times