WATANABE G1000 Series Instruction Manual page 35

Graphic multi-meter, strain gauge/process input
Hide thumbs Also See for G1000 Series:
Table of Contents

Advertisement

・Acquire measured value and finalize with "SET" key.
(Measured value becomes blue simultaneously with finalizing.)
・When the acquired ZERO value falls under the following case, it is
judged an Error and resetting is performed.
Error 1:When the acquired ZERO value is smaller than -4.0mV/V.
Error 2:When the acquired ZERO value is larger than -3.0mV/V.
・Perform resetting by pressing the "▲" key and then pressing the
"ZERO Calibration" key.
6) Equivalent Calibration
・Set the rated output value (mV/V value) for the strain gauge sensor
and the indicated value at the time.
・The setting rage of the rated output value (SPIN value) for the strain
gauge sensor is 0.1 to 3.0mV/V.
・The setting range of SPAN set value (indicated value) is 100 to 99999.
・Select a set value on the set-value select screen and then set in on the
10-key input screen.
a) SPIN value
・Acquire set value and finalize with "SET" key.
(Set value becomes blue simultaneously with finalizing.
SPIN value (Finalized output value)
・When the acquired SPIN value (finalized output value) falls under the
following case, it is judged an error and resetting is performed.
Error 3:When it is smaller than 0.1mV/V value
Error 4:When it is larger than 3.0mV/V value
・To reset, press "C" to set the set value to "0", or press "▲" to return to
the set-value select screen of the above-mentioned 6) and then press
"SPIN" key.
・Set-value select screen
Select the set value (SPIN, SPAN).
SPIN value error
35/118