Itemized setting screen 5/6
Itemized setting screen 6/6
12) Digital ZERO Backup
・ON/OFF setting can be performed by pressing the "Digital ZERO
Backup" key one at a time.
・When backup is set to ON, Digital ZERO continues even when
power is turned OFF/ON.
・It is possible to backup for each channel.
13) Language
・It is possible to change the language displayed by each key
and monitor. Choose between Japanese and English
・Press "Language" key to choose between Japanese and English.
14) Communication and Related Items
・The following setting can be performed on the set-value select
screen.
・BPS : 2400, 4800, 9600, 19200, 38400
・Data Length : 7, 8
・Parity : Even, Odd, None
・Stop Bit : 1, 2
・Delimiter : CR+LF, CR, LF
Caution:Even if address is changed while the line is occupied, it
will change after returning to normal operation once.
15) Address (RS-485 ID)
・Set it on the 10-key input screen.
・Setting range : 0 to 99
・When the output unit is RS-485, the setting screen appears.
Caution : Even if address is changed while the line is occupied, it
will not be adopted but will be adopted when the line is released.