--> Setting example 2(minus mode operation)
Each set value
HH Limit
HI Limit
LO Limit
LL Limit
HH set value (HH Limit) (350)
HI set value (HI Limit) (250)
LO set value (LO Limit) (150)
LL set value (LL Limit) (50)
Output HH
Output HI
Output GO
Output LO
Output LL
2) Area type (Area)
・Comparison set value : setting range ±99999
・Setting conditions : HH set value > HI set value > LO set value > LL set value
・Hysteresis : not used.
3) Rank type (Rank)
・Comparison set value : setting range ±99999
・Hysteresis set value : setting range ±0 to 9999
・Setting conditions:for each comparison set value, hysteresis set value.
Each hysteresis set value
350
HH Hysteresis -15
250
HI Hysteresis -15
150
LO Hysteresis +15
50
LL Hysteresis +15
Indicated value
OFF
ON
OFF
ON
OFF
ON
OFF
ON
OFF
ON
(HH Limit)
Each set value should be set according to setting conditions.
Setting can't be performed if conditions are not satisfied.
Polarity of hysteresis (HH Hysteresis to LL Hysteresis) is automatically set to
"+", operating as plus mode.
・(HH set value) - (+HH Hysteresis) ≧ HI set value
(HH Limit) - (+HH Hysteresis) ≧ HI Limit
・(HI set value) - (+HI Hysteresis) ≧ LO set value
(HI Limit) - (+HI Hysteresis) ≧ LO Limit
・(LO set value) - (+LO Hysteresis) ≧ LL set value
(LO Limit) - (+LO Hysteresis ) ≧ LL Limit
Each set value should be set according to setting conditions.
Setting can't be performed if conditions are not satisfied.
Each hysteresis value
350+15=365
250+15=265
150-15=135
50-15= 35
Indicated value
(HI Limit)
(LO Limit)
49/118
Hysteresis value (365)
Hysteresis value (265)
Hysteresis value (135)
Hysteresis value (35)
(LL Limit)