Wave Start Level
(3)
・When the indicated value satisfies「Wave Start Level + Start Condition」, clear the screen and start
the acquisition of waveform and graphic drawing.
When the acquisition of 2048 points of waveform is finished, hold the display screen.
Wait the next satisfaction of「Wave Start Level + Start Condition」.
・"START" key or START signal is not used.
Example : Start Condition = Case of Rising Edge
Wave start level
(4) E.S. + W.S
・When "START" key or START signal turns ON, and the indicated value satisfies「Wave Start Level +
Start Condition」, clear the screen and start the acquisition of waveform and graphic drawing.
・If "START" key or STOP signal turns ON during the time, cancel the drawing, or if the acquisition of
2048 points of waveform is finished, hold the display screen.
Wait the next "START" key or START signal turning ON.
Example : Start Condition = Case of Pass
Wave start level
"START" key
OFF
ON
START signal OFF
ON
STOP signal
OFF
ON
Start the acquisition of
waveform and graphic drawing
Graphic drawing period (acquire 2048)
Start the acquisition of
waveform and graphic drawing
Graphic drawing period (acquire 2048)
44/118
Hold display screen and
wait for the next
satisfaction of condition
Hold display screen
and wait for the
next START.