WATANABE G1000 Series Instruction Manual page 40

Graphic multi-meter, strain gauge/process input
Hide thumbs Also See for G1000 Series:
Table of Contents

Advertisement

There are 4 different data to be set.
Example of scaling setting
Full Scale Input = 9.000V
Full Scale (indicated value) = 2000
Offset Scale Input = 0.000V
Offset Scale (indicated value) = 0
Example of scaling setting
Full Scale Input = 10.000V
Full Scale (indicated value) = 2000
Offset Scale input = 2.000V
Offset Scale (indicated value) = 0
Example of scaling setting
Full Scale Input = 20.000mA
Full Scale (indicated value) = 2000
Offset Scale Input = 0mA
Offset Scale (indicated value) = 0
Example of scaling setting
Full Scale Input = 18.000mA
Full Scale (indicated value) = 2000
Offset Scale Input= 2.000mA
Offset Scale(indicated value) = 500
・When the "A/D Output Value" key is
pressed, perform on the measured-
value acquisition screen.
・Acquire set value and finalize with
"SET" key. (Measured value becomes
blue simultaneously with finalizing.)
・To reset, press "▲" to return to the
above-mentioned select screen, and
press the "A/D Output Value" key
again.
Display
2000
1500
1000
500
0 2.000
6.000
4.000
Display
2000
1500
1000
500
0
2.000
6.000
-500
4.000
Display
2000
1500
1000
500
0
4.000
12.000
8.000
Display
2000
1500
1000
500
0
4.000
12.000
8.000
Input
9.000
(V)
8.000
Input
10.000
(V)
8.000
Input
20.000
(mA)
16.000
Input
20.000
(mA)
16.000
40/118