WATANABE G1000 Series Instruction Manual page 76

Graphic multi-meter, strain gauge/process input
Hide thumbs Also See for G1000 Series:
Table of Contents

Advertisement

Check result 1
Check result 2
Check result 3
Check result 4
Check result 5
There are check result 1 to 7.
・Check result 1 : All of them display the normal states.
In case of normal, "Normal" is displayed.
・Check result 2 : It is a case where A/D in this Meter is
faulty
When A/D in this Meter is faulty, Error is displayed as the
check result of「A/D」.
When A/D in this Meter is faulty,「+SIG」,「-SIG」, and
「SENSOR」cannot be checked, but the check result of
「+SIG」,「-SIG」, and「SENSOR」is displayed as Normal.
・Check result 3 : It is a case where「+SIG」 line is
disconnected
When connection between the sensor and +SIG terminal is
disconnected, the check result of「+SIG」is displayed as
Error.
In case of disconnection, 「 SENSOR」 cannot be checked, but
the check result of「SENSOR」is displayed as Normal.
・Check result 4
When connection between the sensor and -SIG terminal is
disconnected, the check result of「-SIG」is displayed as
Error.
In case of disconnection, 「 SENSOR」 cannot be checked, but
the check result of「SENSOR」is displayed as Normal.
・Check result 5
When connection between the sensor, +SIG and -SIG
terminal is disconnected, the check result of「+SIG」and
「-SIG」is displayed as Error.
In case of disconnection,「SENSOR」cannot be checked, but
the check result of「SENSOR」is displayed as Normal.
76/118