WATANABE G1000 Series Instruction Manual page 31

Graphic multi-meter, strain gauge/process input
Hide thumbs Also See for G1000 Series:
Table of Contents

Advertisement

Example of ZT display
ZT
Yes
Zero
Display
display
Indicated value 0
Indicated value "0"
Itemized setting screen 3/6
Yes
No
No
Zero
Normal
Normal
display
display
display
ZT cycle
Normal display xx
6) Input ON Delay
・Set time on the 10-key input screen.
Time-setting range (Input ON Delay) : 0 to 4.999 seconds.
・This function is provided to adjust the duration from the time
when the control terminal and "HOLD" key are activated to the
time when operation is actually started.
STA, STB OFF ("1")
ON ("0")
"hold"key OFF ("1")
ON ("0")
Internal
OFF ("1")
operation
ON ("0")
・Use Input on delay for the above-mentioned control terminal and
"HOLD" key.
・STA (Ach) / STB (Bch) control terminals can be used separately for
each channel.
・"HOLD" key : Operates at Ach and Bch simultaneously.
・Time setting is commonly used for Ach/Bch.
7) Output OFF Delay (Output Delay)
(Not used for Waveform Comparison / Waveform & Displacement
Comparison)
・Set time on the 10-key input screen.
Delay range (Output Delay) : 0 to 4.999 seconds.
・Used for comparison output (OUT1 to OUT5) and synchronizing
signal output (SYNC).
・Set the time required from turning OFF of comparison output and
synchronizing signal to actual turning OFF.
・When comparison output is turned ON and OFF during output off
delay time, operation is repeated from the start of time setting.
Yes
Yes
Normal
Zero
display
display
ZT Width
ZT Width
*1 Indicated value =
"0" in this range.
Indicated value "0"
Input on delay
31/118
*1

Advertisement

Table of Contents
loading

Table of Contents